首页 | 本学科首页   官方微博 | 高级检索  
     


Amperometric Determination of Cr(III) and Cr(VI) by Flow Injection Analysis
Authors:J.M. Hwang  Y.M. Chen
Abstract:Speciation of Cr(III) and Cr(VI) can be attained by flow injection analysis with amperometric detection. Cr(VI) is reduced in an acidic medium to Cr(III) with a glassy carbon electrode at —0.1 V vs. Ag/AgCl and the current is recorded. Cr(III) is oxidised on-line to Cr(VI) with alkaline hydrogen peroxide solution. From the difference of the total chromium and Cr(VI), the amount of Cr(III) was obtained. A linear calibration curve for Cr(VI) was obtained for the concentration ranges 0.01-5.0ppm of Cr(VI) and we have calculated the limit of determination to be about 0.5ppb. We have studied the degree of reproducibility obtained using the solid electrodes under various conditions. The influence of flow rate, coil length, interfenences and the extent of reaction were studied.
Keywords:Flow injection analysis  speciation of Cr(III)  Cr(VI)   amperometric detection
设为首页 | 免责声明 | 关于勤云 | 加入收藏