首页 | 本学科首页   官方微博 | 高级检索  
     


Analysis of opioid peptides by on-line SPE-CE-ESI-MS
Authors:Hernández Elena  Benavente Fernando  Sanz-Nebot Victoria  Barbosa José
Affiliation:Department of Analytical Chemistry, University of Barcelona, Barcelona, Spain.
Abstract:In this study, SPE-CE-ESI-MS is explored for the preconcentration and separation of dilute solutions of six opioid peptides. First, a CE-ESI-MS methodology was developed and validated. LODs of around 1 microg/mL were obtained for all the studied peptides. For SPE-CE-ESI-MS experiments, a home-made SPE microcartridge containing a C18 sorbent was constructed near the inlet of the separation capillary. After optimizing the on-line preconcentration methodology, LODs between 10 and 0.1 ng/mL were achieved. Repeatability, reproducibility, durability of the microcartridges and linearity of the SPE-CE-ESI-MS methodology were also investigated and compared to the values obtained by CE-ESI-MS. Finally, human plasma samples fortified with opioid peptides were analyzed by SPE-CE-ESI-MS in order to show the potential of the methodology for the analysis of biological fluids.
Keywords:CE  MS  Neuropeptides  On‐line preconcentration
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号