Effects of sampler-skimmer separation in inductively coupled plasma-mass spectrometry
Authors:J. W. H. Lam and G. Horlick
Affiliation:

Department of Chemistry, University of Alberta, Edmonton, Alberta, Canada T6G 2G2

Abstract:The effects of varying the sampler-skimmer spacing in an inductively coupled plasma-mass spectrometer are illustrated. The signals for a number of species were monitored including background ions (ArN+, ArO+, ArOH+, Ar2+, Ar2H+), background continuum, analyte ions (Cu+, Ce+, La+), analyte oxide ions (CeO+, LaO+), and matrix induced ions (ArNa+). As the sampler-skimmer spacing is increased over that normally used, the signal for some species decreases in intensity while the signal for others increases in intensity. There is a wide range of differential behavior and in some cases the signal changes are quite dramatic.

In addition to presenting results for an Ar plasma, measurements are also presented for a N2-Ar mixed gas plasma. For some species, the signal changes observed when the sampler-skimmer spacing is increased are quite different with the N2-Ar mixed gas plasma than with the Ar plasma and are, in some cases, opposite to signal changes observed with the Ar plasma. It appears that a larger sampler-skimmer spacing is more appropriate for a N2 mixed gas plasma than the spacing normally used for a pure Ar plasma. Finally the effect of sampler-skimmer spacing on matrix effects is presented and it appears that matrix effects are not induced in the zone between the sampler and the skimmer.