首页 | 本学科首页   官方微博 | 高级检索  
     


Quantitation of insulin injection by high-performance liquid chromatography and high-performance capillary electrophoresis.
Authors:M Lookabaugh  M Biswas  I S Krull
Affiliation:US Food and Drug Administration, Winchester Engineering and Analytical Center, MA 01890.
Abstract:High-performance capillary electrophoresis (HPCE) was evaluated as a potential technique for the regulatory analysis of commercial dosage forms of insulin. A comparison was made to a liquid chromatographic analysis presently being proposed as an official monograph in the United States Pharmacopeia. The salient points of this comparison were accuracy, precision and ease of use. Both authentic (i.e. single blind, spiked) samples and commercial pharmaceutical formulations (injections) were examined. Chromatographic analyses of both commercial formulations and authentic samples were characterized by good precision, with accuracy being supported by results from authentic (spiked) samples. Conventional HPCE (by which is meant a non-micellar electrolyte used with an uncoated, unmodified fused-silica capillary) achieved reasonable accuracy, but less than impressive precision, when applied to authentic samples. When used for commercial formulations, this type of HPCE did not produce a level of accuracy suitable for regulatory purposes, even with the use of an internal standard.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号