首页 | 本学科首页   官方微博 | 高级检索  
     


Highly sensitive optical sensor system for blood leakage detection
Authors:Masahiro Ueda   Kazuhiko Ishikawa   Chen Jie  Sanae MizunoYasunori Touma
Affiliation:

Faculty of Education, Fukui University, Bunkyo 3-9-1, Fukui 910, Japan

Hokkei Industry Co. Ltd, Masuizumi 3-4-20, Kanazawa 921, Japan

Abstract:A highly sensitive method for the detection of blood leakage has been developed, and a practical sensor system for blood concentration measurement has been constructed. The present method is based on the attenuation of laser light by blood cells. The effects of the fluctuations of the incident laser light power are eliminated by normalizing the attenuated light intensity by the incident light intensity. A part of the incident laser light is reflected by a beam splitter mounted at the entrance of the test cell, of which the power is measured to provide base data for normalization. The optical path is extended to enhance sensitivity by using a pair of side mirrors. This multi-reflection method is very effective to increase sensitivity; the maximum sensitivity obtained for blood concentration is about 4 X 10-6 by volume, which is significantly higher than that of the conventional sensors.
Keywords:
本文献已被 ScienceDirect 等数据库收录!