Frequency response correction of hot-film measurements in a turbulent airflow |
| |
Authors: | C Ljus B Johansson A -E Almstedt |
| |
Institution: | Department of Thermo and Fluid Dynamics Chalmers University of Technology S-412 96 G?teborg, Sweden, SE
|
| |
Abstract: | Turbulence measurements were performed in a fully developed turbulent pipe flow using a wedged-shaped hot-film probe. The
dynamic sensitivity of hot-film probes is frequency dependent and cannot be obtained from static calibration. As it is complicated
to make a dynamic calibration, a correction method for the turbulent energy spectrum is used instead. The measurements were
compared with hot-wire turbulence measurements and show good agreement.