首页 | 本学科首页   官方微博 | 高级检索  
     

Zeeman-石墨炉原子吸收测定全血中微量铅
引用本文:陈晓红,于笑宇. Zeeman-石墨炉原子吸收测定全血中微量铅[J]. 光谱学与光谱分析, 2005, 25(3): 477-478
作者姓名:陈晓红  于笑宇
作者单位:宁波市疾病预防控制中心,浙江 宁波 315010
摘    要:提出用0.5%硝酸稀释样品,Zeeman-石墨炉原子吸收光谱法直接测定全血中微量铅,并讨论了有关的实验条件。方法的检出限为10.0 pg,标样的相对标准偏差(RSD)0.36%~1.8%,样品的相对标准偏差(RSD)1.4%~2.3%,回收率96.0%~101.7%,共存物干扰小。利用此法已分析600多例铅接触工人和儿童,结果满意。本法简单,快速是一种理想的全血测定方法。

关 键 词:Zeeman-石墨炉原子吸收  全血    
文章编号:1000-0593(2005)03-0477-02
收稿时间:2003-08-16
修稿时间:2003-08-16

Determination of Trace Pb in Whole Blood by Zeeman-GFAAS
CHEN Xiao-hong,YU Xiao-yu. Determination of Trace Pb in Whole Blood by Zeeman-GFAAS[J]. Spectroscopy and Spectral Analysis, 2005, 25(3): 477-478
Authors:CHEN Xiao-hong  YU Xiao-yu
Affiliation:Department of Laboratory Test, Ningbo Municipal Center for Disease Prevention and Control, Ningbo 315010, China
Abstract:The determination method for trace Pb in whole blood was studied, which was diluted by 0.5% HNO_3. In the measurement, Zeeman-GFAAS was used and the relative condition was discussed. This method shows that the detection limit was 10.0 pg, the relative standard deviation of standards (RSD) was 0.36%-1.8%, the relative standard deviation of samples (RSD) was 1.4%-2.3%, and the recovery rate was 96.0%-101.7%. The concurrent interference was low. This method is simple and rapid. It is an ideal method to determine trace Pb in whole blood.
Keywords:Zeeman-GFAAS  Whole blood  Pb