首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of plutonium and uranium in mixed oxide fuels by sequential redox titration
Authors:Chadwick P H  McGowan I R
Institution:British Nuclear Fuels Limited, Windscale and Calder Works, Sellafield, Seascale, Cumberland, U.K.
Abstract:The use of a sequential determination of uranium and plutonium in a single sample solution results in a saving in analysis time and apparatus requirements. The method starts with U(IV) and Pu(in) in a mixture of sulphuric and nitric adds. Titration with dichromate, using amperometry at a pair of polarizable electrodes, produces two well-defined end-points corresponding to the sequential oxidation of U(IV) to U(VI) and Pu(III) to Pu(IV). The quantitative oxidation of U(IV) to U(VI) is achieved via the action of Pu(IV) as intermediate, and is dependent upon establishing conditions which favour rapid reaction between U(IV) and Pu(IV). The method is precise and accurate. With Pu-U mixtures containing between 15 and 30% plutonium the precision (3sigma) of the Pu: U ratio results is +/-0.6% on samples containing 100-120 mg of plutonium plus uranium. Iron and vanadium interfere quantitatively with plutonium, copper interferes non-quantitatively with uranium, and gross amounts of molybdenum mask the uranium end-point.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号