首页 | 本学科首页   官方微博 | 高级检索  
     


Forced-flow liquid chromatography with a coulometric detector
Authors:Johnson D C  Larochelle J
Affiliation:Department of Chemistry, Iowa State University, Ames, Iowa 50010, U.S.A.
Abstract:A tubular platinum electrode packed with small chips of platinum is described which was shown to function with 100% electrolytic efficiency. Application of the electrode is described for the coulometric determination of electroactive species in the effluent from a liquid chromatograph. Cu and Fe were determined in a series of standards. An average relative deviation of less than 2 ppt was obtained for analysis of samples containing 5 mug of Cu(II) and 5 mug of Fe(III).
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号