首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Tritium monitoring with a scintillation chamber
Authors:M Chudý  R Janik  P Povinec
Institution:(1) Department of Nuclear Physics, Comenius University, Mlynská dolina, Bratislava, (Czechoslovakia)
Abstract:A tritium monitor based on plastic scintillation sheets is described. The sensitive volume of the scintillation chamber is 0.95 l. The chamber with coincidence electronics has a discrimination capability and permits to detect tritium in the presence of other radionuclides and external gamma-radiation. The monitor characteristics obtained with tritiated air are presented. The monitor sensitivity for tritium in air is 5.6 cpm for each kBq·m−3. The monitoring of tritium in air and water is possible using the instrument described.