Measurement of catecholamines, their precursor and metabolites in human urine and plasma by solid-phase extraction followed by high-performance liquid chromatography with fluorescence derivatization |
| |
Authors: | H Nohta E Yamaguchi Y Ohkura H Watanabe |
| |
Affiliation: | Faculty of Pharmaceutical Sciences, Kyushu University, Fukuoka Japan. |
| |
Abstract: | A high-performance liquid chromatographic method is described for the determination in human urine and plasma of catecholamines, their precursor and metabolites [amino compounds (norepinephrine, epinephrine, dopamine, normetanephrine, metanephrine, 3-methoxytyramine and L-DOPA), acidic compounds (3,4-dihydroxymandelic acid, 3,4-dihydroxyphenylacetic acid, vanillylmandelic acid and homovanillic acid) and alcoholic compounds (3,4-dihydroxyphenylethyleneglycol and 4-hydroxy-3-methoxyphenylethyleneglycol)]. Urine (0.5 ml) containing 3,4-dihydroxybenzylamine and 4-hydroxy-3-methoxycinnamic acid (internal standards) is deproteinized with perchloric acid, and the resulting solution is fractionated by solid-phase extraction on a strong cation-exchange resin cartridge (Toyopak IC-SP S) into two fractions (amine fraction and acid-alcohol fraction), which include 3,4-dihydroxybenzylamine and 4-hydroxy-3-methoxycinnamic acid, respectively. Plasma (0.7 ml) is deproteinized in the presence of 3,4-dihydroxybenzylamine (internal standard) in the same manner, and the resulting solution is directly used as an acid-alcohol fraction, while an amine fraction is obtained as for urine. Each fraction is subjected to the previously established ion-pair reversed-phase chromatography with post-column derivatization involving coulometric oxidation followed by fluorescence reaction with 1,2-diphenylethylenediamine. The detection limits, at a signal-to-noise ratio of 5, of the compounds measured in urine are 300 pmol/ml for the two mandelic acids, 2-7 pmol/ml for the other acidic and alcoholic compounds, 12 pmol/ml for L-DOPA and 0.6-2 pmol/ml for the other amino compounds; the corresponding values for plasma samples are 80, 0.5-3, 10 and 0.6-3 pmol/ml, respectively. |
| |
Keywords: | |
|
|