首页 | 本学科首页   官方微博 | 高级检索  
     


Correction of intensity nonuniformity in spin-echo T(1)-weighted images
Authors:Collewet G  Davenel A  Toussaint C  Akoka S
Affiliation:Cemagref, 17 Avenue de Cucillé, 35044 Rennes Cedex, France. guylaine.collewet@cemagref.fr
Abstract:This paper presents a method to correct intensity nonuniformity in spin-echo T(1)-weighted images and particularly the inhomogeneities due to RF transmission imperfections which have tissue-dependent effects through the T(1) relaxation times. This method is based on the use of a uniform phantom, first for classic normalization by division by the phantom images, and second for T(1)-correction using the RF transmitted cartography. We present experimental results from a bi-phasic (oil/water) phantom and from a salmon with a 0.2 T imager. The results demonstrate the efficiency of the method in the two cases and its ability to cope with partial volume effect.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!