Construction and Evaluation of a Promazinium Cation-Selective Electrode |
| |
Authors: | Magda M. Ayad Abdalla Shalaby Hisham E. Abdellatef Heba M. Elsaid |
| |
Affiliation: | (1) Analytical Chemistry Department, Faculty of Pharmacy, Zagazig University, Zagazig, Egypt, EG |
| |
Abstract: | The construction of a plasticised PVC matrix-type promazinium cation-selective membrane electrode and its use in the potentiometric determination of promazine hydrochloride in pharmaceutical preparations are described. It is based on the use of the ion-associate species, formed by promazinium cation and tetraphenylborate (TPB) counter ion. The basic electrode performance characteristics are evaluated according to IUPAC recommendations. It exhibited a linear response for 1 × 10−2−1 × 10−5 M of promazine hydrochloride solutions with a cationic Nernstian slope over the pH range 2–6. Common organic and inorganic cations showed negligible interference. Direct potentiometric determination of 1 × 10−2−1 × 10−5 M aqueous promazine hydrochloride using this membrane electrode system showed an average recovery of 99.5% with a mean standard deviation of 1.5%. This electrode was successfully used for monitoring the titration of promazine hydrochloride with sodium tetraphenyl borate and for determining promazine hydrochloride in ampoules. Received June 15, 2001 Revision November 6, 2001 |
| |
Keywords: | : Promazine hydrochloride ion-selective electrode tetraphenylborate potentiometric titration pharmaceutical analysis. |
本文献已被 SpringerLink 等数据库收录! |