首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Preconcentration of trace amounts of uranium in water samples on octadecyl silica membrane disks modified by bis(2-ethylhexyl) hydrogen phosphate and its determination by alpha-spectrometry without electrodeposition
Authors:Y Yamini  P Ashtari  A R Khanchi  M Ghannadi-Maragheh  M Shamsipur
Institution:(1) Department of Chemistry, Tarbiat Modarres University, Tehran, Iran;(2) Jaber Ibn Hayan Research Laboratories, Atomic Energy Organization of Iran, Tehran, Iran;(3) Department of Chemistry, Razi University, Kermanshah, Iran
Abstract:A simple and reliable method for the selective extraction and determination of uranium in water using octadecyl-bonded silica membrane disks modified with bis (2-ethylhexyl) hydrogen phosphate and alpha-spectrometry is described. Extraction efficiency and influence of sample matrix, optimum amount of extractant, type and minimum amount of organic eluent and flow rates were evaluated. The limit of detection of the proposed method is 40 ng per 1000 ml. The influence of potential interfering cations in water samples on the recovery of U(VI) was investigated. The method was successfully applied to the extraction and determination of uranium in natural water.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏