首页 | 本学科首页   官方微博 | 高级检索  
     


An improved flow-based procedure for microdetermination of total tannins in beverages with minimized reagent consumption
Authors:Carlos M. C. Infante  Vitória R. B. Soares  Mauro Korn  Fábio R. P. Rocha
Affiliation:Instituto de Química, Universidade de S?o Paulo, S?o Paulo, SP, Brazil
Departamento de Ciências Exatas e da Terra, Universidade do Estado da Bahia, Salvador, BA, Brazil
Abstract:A flow system designed with solenoid micro-pumps is introduced for spectrophotometric determination of total tannins based on the Folin– Denis reaction. The procedure minimizes the main drawbacks related to the AOAC batch procedure, i.e. interferences from reducing species in the samples, high reagent consumption and waste generation, and low sampling rate. Linear response was observed for tannic acid concentrations in the range 2–100 mg L−1, with a detection limit (99.7% confidence level) of 0.3 mg L−1. The sampling rate and coefficient of variation (n = 10) were estimated as 75 measurements per hour and 1.1%, respectively. Results of determination of total tannin in tea, beer and wine samples were in agreement with those achieved by the batch reference procedure at the 95% confidence level. In comparison to the batch procedure, the reagent consumption and effluent generation were 83 and 60-fold lower, respectively. Correspondence: Fábio R. P. Rocha, Instituto de Química, Universidade de S?o Paulo, PO Box 26077, S?o Paulo, 05513-970 SP, Brazil
Keywords:: Flow analysis   multicommutation   multi-pumping   tannin   Folin–  Denis reagent
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏