首页 | 本学科首页   官方微博 | 高级检索  
     


Capillary electrophoresis with inductively coupled plasma-mass spectrometric and electrospray time of flight mass spectrometric detection for the determination of arsenic species in fish samples
Authors:Meermann Björn  Bartel Marc  Scheffer Andy  Trümpler Stefan  Karst Uwe
Affiliation:Institute of Inorganic and Analytical Chemistry, University of Münster, Münster, Germany.
Abstract:CE was coupled to inductively coupled plasma MS (ICP-MS) and ESI-MS to identify and quantify the arsenic species arsenobetaine (AsB), arsenite (As(III)), arsenate (As(V)), and dimethylarsinic acid (DMA). A GC-flame ionization detector (FID)-based German standard method and ICP-MS were used for validation of the data obtained for arsenobetaine and total arsenic, respectively. LODs obtained with the CE-ESI-TOF-MS method were 1.0x10(-7) M for AsB, 5.0x10(-7) M for DMA, and 1.0x10(-6) M for As(III) and As(V). For the CE-ICP-MS method, LODs were 8.5x10(-8) M for AsB, 9.5x10(-8) M for DMA, 9.3x10(-8) M for As(III), and 6.2x10(-8) M for As(V). While CE-ICP-MS provided high sensitivity and better reproducibility for quantitative measurements, CE-ESI-MS with a TOF mass analyzer proved to be valuable for species identification. With this setup, fish samples were prepared and analyzed and the obtained data were successfully validated with the independent methods.