首页 | 本学科首页   官方微博 | 高级检索  
     


Signal stability of Nafion-coated thin mercury film electrodes for stripping voltammetry
Authors:Hoyer B  Jensen N
Affiliation:Department of Chemistry, Aarhus University, Langelandsgade 140, 8000 Arhus C, Denmark.
Abstract:The signal stability of the Nafion-coated thin mercury film electrode (NCTMFE) was studied by using cadmium and lead as test analytes and differential pulse anodic stripping voltammetry as detection method. In particular, the effect of the casting solvent and the curing procedure employed in the preparation of the polymer film was examined. Best results were obtained with N,N-dimethylacetamide as casting solvent and a two-step curing procedure in which the polymer was evaporated to dryness at 55 degrees and cured at 105 degrees with a hot-air gun. Mercury plating was performed ex situ. An NCTMFE prepared in this manner has a better signal stability than ex situ-plated as well as in situ-plated conventional mercury film electrodes.
Keywords: