首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Trace determination of sulphide and sulphur dioxide by vapor molecular absorption spectrometry using magnesium and tellurium hollow cathode lamps
Authors:Qinhan Jin  Hanqi Zhang  Yixiang Duan  Aimin Yu  Xinwei Liu  Lishuang Wang
Institution:Department of Chemistry, Jilin University, Changchun, 130023, People's Republic of China.
Abstract:A new method has been developed in this paper for determining sulphide and sulphur dioxide simultaneously. Two emission lines, 202.6 nm and 214.3 nm which are emitted from hollow cathode lamps (HCL) of magnesium and tellurium, respectively, were used as radiation sources for measurement of absorbances of H(2)S at 202.6 nm and SO(2) at 214.3 nm or 202.6 nm. The detection limit for S(2-) was shown to be 0.01 mug/ml and the detection limits for SO(2) with 202.6 nm and 214.3 nm lines were 0.05 and 0.2 mug/ml, respectively. The method has been employed to satisfactorily analyse practical samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏