首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Analogue implementation of analytic signal processing for pulse-echo systems
Authors:PM Gammell  
Institution:

a Information Systems Research Section, California Institute of Technology, Jet Propulsion Laboratory, Pasadena, California 91109, USA

b Radiation Physics Section, Department of Radiology, University of California Medical School, Los Angeles, California 90033, USA

Abstract:An alternative to rectification is proposed for detection of an ultrasonic signal. This method is especially useful in medical and non-destructive evaluation (nde) applications. With this method, the magnitude of the complex analytic signal is used to define the envelope of the ultrasonic waveform. The square of this quantity has been shown elsewhere to be equal to the true rate-of-arrival of energy. An earlier study, using digital data processing, has already demonstrated the superior resolvability of closely spaced interfaces obtained with the analytic signal magnitude, as compared to conventional rectification. Here, an analogue implementation is presented which utilizes single-sideband techniques to obtain both quadrature components of the analytic signal and its magnitude. A conventional transducer, pulser, and receiver are used.
Keywords: