首页 | 本学科首页   官方微博 | 高级检索  
     


Potentiometric sensors based on organic ion exchangers for determining tetraalkylammonium salts
Authors:E. G. Kulapina  N. M. Makarova  L. A. Bazhanova  E. S. Pogorelova
Affiliation:1. Saratov State University, ul. Astrakhanskaya 8, Saratov, 410012, Russia
Abstract:Potentiometric sensors with plasticized polymer membranes based on organic ion exchangers, tetraalkylammonium dodecyl sulfates (benzyldimethyldodecylammonium, benzyldimethyltetradecylammonium, dimethyldistearylammonium), have been proposed for the determination of quaternary ammonium salts in model solutions and KATAPAV technical solutions. The thermal stability, composition, and solubility product have been estimated. It has been shown that ion associates are stable to 60?C70°C, K S varies in the range from 2 × 10?8 to 5 × 10?10. The basic electrochemical parameters of the sensors have been determined as well, such as linearity ranges of the electrode function (5 × 10?5 (5 × 10?6)?1 × 10?2 (1 × 10?3) M) and slopes of the electrode functions (47?C59 mV/pc), response time (60?C90 sec), potential drift (2?C3 mV/day), operation period (3?C4 months), limits of detection for tetramethylammonium salts (1 × 10?5?4 × 10?7 M).
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏