首页 | 本学科首页   官方微博 | 高级检索  
     检索      

便携式红外热像仪外场测温误差分析及修正
引用本文:王吉军,周校,李岩,李芳媛,余松林.便携式红外热像仪外场测温误差分析及修正[J].红外,2021,42(10):28-32.
作者姓名:王吉军  周校  李岩  李芳媛  余松林
作者单位:军事科学院国防工程研究院,北京100039
基金项目:装备发展部装备试验鉴定技术研究项目
摘    要:红外测温成像技术精度的影响因素较多。尤其在实验外场的复杂环境下,有时难以达到预期指标。为了有效提高便携式红外热像仪的外场测温精度,利用红外测温理论模型梳理和总结了外场测温的主要影响因子,然后利用外场实测数据对其进行了统计分析,并提出了相应的修正方法。结果表明,修正后的数据标准差由0.89 K降至0.3 K,有效提高了测温精度。

关 键 词:红外成像  非接触式  误差影响  测量精度
收稿时间:2021/5/27 0:00:00
修稿时间:2021/6/9 0:00:00

Error Analysis and Correction of Field Temperature Measurement for Portable Infrared Thermal Imager
wangjijun,zhou xiao,liyan,lifangyuan and yusonglin.Error Analysis and Correction of Field Temperature Measurement for Portable Infrared Thermal Imager[J].Infrared,2021,42(10):28-32.
Authors:wangjijun  zhou xiao  liyan  lifangyuan and yusonglin
Institution:Institute of national defense engineering, Academy of Military Sciences,Institute of national defense engineering,Academy of Military Sciences,Institute of national defense engineering, Academy of Military Sciences,Institute of national defense engineering, Academy of Military Sciences,Institute of national defense engineering, Academy of Military Sciences
Abstract:There are many factors affecting the accuracy of infrared temperature imaging technology. Especially in the complex environment of the experimental field, sometimes it is difficult to reach the expected index. In order to effectively improve the measurement accuracy of portable infrared thermal imager in the field, the main influencing factors of the field temperature measurement are sorted and summarized by the infrared temperature measurement theoretical model, and the influence factors are statistically analyzed by using the measured data in the field in the paper. In addition, the corresponding correction method is proposed. The results show that the standard deviation is reduced from 0.89 K to 0.3 K after the correction, which effectively improves the temperature measurement accuracy.
Keywords:infrared imaging  non-contact  error influence  measurement accuracy
本文献已被 万方数据 等数据库收录!
点击此处可从《红外》浏览原始摘要信息
点击此处可从《红外》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号