首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Precise controlled-potential coulometry: Instrumental errors
Authors:A N Mogilevskii
Institution:(1) Vernadsky Institute of Geochemistry and Analytical Chemistry, Russian Academy of Sciences, ul. Kosygina 19, 117975 Moscow, Russia
Abstract:Instrumental errors in modern controlled-potential coulometry are analyzed. The instrumental errors are classified into two main groups, namely, errors of determining the degree of completion of analyte electrolysis and errors of measuring the quantity of electricity taken for electrolysis. Various components of these errors have been estimated numerically. It has been demonstrated that, with the use of currently available electronic components, the total instrumental error can be as low as 0.01–0.02%, which allows the analysis error to be reduced down to 0.05–0.1%. Presented at the V All-Russian Conference with the Participation of CIS Countries on Electrochemical Methods of Analysis (EMA-99), Moscow, December 6–8, 1999.
Keywords:
本文献已被 SpringerLink 等数据库收录!