Rapid extraction and quantitative detection of the herbicide diuron in surface water by a hapten-functionalized carbon nanotubes based electrochemical analyzer |
| |
Authors: | Sharma Priyanka Bhalla Vijayender Tuteja Satish Kukkar Manil Suri C Raman |
| |
Institution: | Institute of Microbial Technology (CSIR), Sector 39-A, Chandigarh 160036, India. |
| |
Abstract: | A solid phase extraction micro-cartridge containing a non-polar polystyrene absorbent matrix was coupled with an electrochemical immunoassay analyzer (EIA) and used for the ultra-sensitive detection of the phenyl urea herbicide diuron in real samples. The EIA was fabricated by using carboxylated carbon nanotubes (CNTs) functionalized with a hapten molecule (an amine functionalized diuron derivative). Screen printed electrodes (SPE) were modified with these haptenized CNTs and specific in-house generated anti diuron antibodies were used for bio-interface development. The immunodetection was realized in a competitive electrochemical immunoassay format using alkaline phosphatase labeled secondary anti-IgG antibody. The addition of 1-naphthyl phosphate substrate resulted in the production of an electrochemically active product, 1-naphthol, which was monitored by using differential pulse voltammetry (DPV). The assay exhibited excellent sensitivity and specificity having a dynamic response range of 0.01 pg mL(-1) to 10 μg mL(-1) for diuron with a limit of detection of around 0.1 pg mL(-1) (n = 3) in standard water samples. The micro-cartridge coupled hapten-CNTs modified SPE provided an effective and efficient electrochemical immunoassay for the real-time monitoring of pesticides samples with a very high degree of sensitivity. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|