Uniformity of ionization response of structurally diverse analytes using a chip-based nanoelectrospray ionization source |
| |
Authors: | Hop Cornelis E C A Chen Yue Yu Li J |
| |
Affiliation: | Pfizer Global Research and Development, Department of Pharmacokinetics, Dynamics and Metabolism, Eastern Point Road, Groton, CT 06340, USA. cornelis.e.hop@pfizer.com |
| |
Abstract: | The major drawback of liquid chromatography/mass spectrometry (LC/MS) for the analysis of mixtures is the non-quantitative nature of these studies. The ionization efficiency of the various components in the mixture (e.g., a compound and its metabolites) can vary greatly and, therefore, relative intensities of signals cannot be related to relative abundance. A chip-based nanoelectrospray ionization source was used to compare the ionization efficiencies of compounds with different physical-chemical characteristics. The data indicate that the ionization efficiencies vary much less with the chip-based device than by LC/MS. This was ascribed to the generation of a much higher electric field around the nozzles, which supplies a large excess of protons to the small droplets and reduces/eliminates the differences in the ionization efficiency for the analytes. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|