Quantitative determination of ochratoxin A by liquid chromatography/electrospray tandem mass spectrometry |
| |
Authors: | Lau B P Scott P M Lewis D A Kanhere S R |
| |
Affiliation: | Health Canada, Health Protection Branch, Food Research Division, Address Locator 2203D, Tunney's Pasture, Ottawa, Ontario, Canada K1A 0L2. |
| |
Abstract: | Mass spectrometry of ochratoxin A (OTA) and B (OTB) under electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) was studied. ESI offers higher sensitivities and less fragmentation than APCI. A sensitive LC/MS/MS method for the determination of ochratoxin A (OTA) in human plasma samples was developed. The absolute minimum detection limit was around 10-20 pg per injection, corresponding to 0.5 ppb in an injection equivalent to 20-40microg of human plasma. Ochratoxin B (OTB) was used as an internal standard and its absence in real-life samples was carefully checked before samples were spiked with the internal standard. It was found that these two ochratoxins are susceptible to sodium adduct formation. Fragment ions from the [M + H](+) and [M + Na](+) ions of both OTA and OTB were monitored in the multiple reaction monitoring mode. Three quantitative approaches, standard addition method, internal standard method (using ochratoxin B as an internal standard) and external standard method, were compared in the analysis of human blood plasma. Results from the mass spectrometric method were comparable to those from a conventional LC/fluorescence method. The LC/MS/MS method was also applied to the analysis of contaminated coffee samples. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|