首页 | 本学科首页   官方微博 | 高级检索  
     


Improved preparative electrochromatography column design
Authors:Kenneth D. Cole  Heriberto Cabezas Jr.
Affiliation:

Biotechnology Division, National Institute of Standards and Technology, Gaithersburg, MD 20899, USA

Abstract:Improved chromatography column fittings were developed for the efficient and reliable application of an electric field to a preparative chromatography column (a process termed electrochromatography). The improved fittings contained electrodes in close proximity to the column packing media and allowed uniform electric fields to be applied. Membranes in the fittings prevented mixing of the electrode and the column eluent buffers. The membranes prevented gases and electrolytic products generated in the electrode chamber from entering the column eluent buffer. An electrode buffer solution was pumped through the electrode chamber to a large external container. The circulation of buffer through the electrode chamber removed the gases and electrolytic products and ensured a uniform electric field by helping to maintain a constant buffer composition. The membranes prevented macromolecules being separated on the column from coming in contact with the electrodes.
Keywords:Electrochromatography   Preparative chromatography   Column design   Electrodes   Membranes   Instrumentation