首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Calibration of ear canals for audiometry at high frequencies
Authors:K N Stevens  R Berkovitz  G Kidd  D M Green
Abstract:A procedure is described for determining the absolute sound pressure at the inner end of the ear canal when a sound source is coupled to the ear, for frequencies in the range 8-20 kHz. The transducer that generates the sound is coupled to the ear canal through a lossy tube, yielding a source impedance that is approximately matched to the characteristic impedance of the ear canal. A small microphone is located in the coupling tube close to the entrance to the ear canal. Calibration is carried out by measuring the response at this microphone when an impulse is applied at the transducer. To estimate the sound pressure at the medial end of the ear canal, the Fourier transform of this impulse response is corrected by an all-pole function in which the poles are estimated from the minima in this Fourier transform. Data on individual ear canals are presented in terms of gain functions relating the sound pressure at the medial end of the ear canal to the sound pressure when the coupling tube is blocked. The average gain function for a group of adult ears increases from 2 to 12 dB over the frequency range 8-20 kHz, in rough agreement with data from ear-canal models. Possible sources of error in the calibration procedure are discussed.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号