Neutron spectrum and doses in a 18 MV LINAC
Authors:Hector Rene Vega-Carrillo  Berenice Hernández-Almaraz  Victor Martin Hernández-Dávila and Arturo Ortíz-Hernández
Institution:(1) Unidad Academica de Estudios Nucleares de la Universidad Autonoma de Zacatecas, Apdo. Postal 336, 98000 Zacatecas, Zac, Mexico
Abstract:Bonner sphere spectrometer with TLDs pairs has been utilized to measure the neutron spectrum 100 cm from isocenter of a 18 MV LINAC, simultaneously the ambient dose equivalent due to neutrons and photons was measured in the control room area with neutron and gamma-ray area monitors. Measurements were carried out when the LINAC was delivering a dose of 600 MU at the isocentre that was located at 5 cm depth of a head phantom. Undesired neutron field in the treatment room produce activation reactions with nuclei in different materials of LINAC, couch, air, and phantom. To determine the dose due to decay of activation nuclei the ambient dose due to gamma-rays was measured inside the treatment room immediately after dose was delivered. Measured spectrum has two peaks, one between 0.1–1 MeV and other in the thermal region, the ambient dose equivalent in the control room are 3.1 and 0.93 μSv h−1 for photons and neutrons, respectively. In the treatment room the ambient dose equivalent due to photons produced during decay of activation nuclei varies from 6 to 26.1 μSv h−1.
Keywords:
本文献已被 SpringerLink 等数据库收录!