Institution:1. Proteome Center Rostock, University Medicine Rostock, Rostock, Germany;2. School of Pharmacy, University of Ghana, Legon, Ghana;2. Department of Cerebrovascular Disease, Kyushu Rosai Hospital, Japan;3. Department of Neurosurgery, Kyushu University, Japan
Abstract:This study describes the application of two in-house developed dosimeters, the Dose Magnifying Glass (DMG) and the MOSkin dosimeter at the Centre for Medical Radiation Physics, University of Wollongong, Australia, for the measurement of CT dose profiles for a clinical diagnostic 16-slice MSCT scanner. Two scanner modes were used; axial mode and helical mode, and the effect of varying beam collimation and pitch was studied. With an increase in beam collimation in axial mode and an increase of CT pitch in helical mode, cumulative point dose at scanner isocentre decreased while FWHM increased. There was generally good agreement to within 3% between the acquired dose profiles obtained by the DMG and the film except at dose profile tails, where film over-responded by up to 30% due to its intrinsic depth dose dependence at low doses.