首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Investigation of microwave assisted drying of samples and evaporation of aqueous solutions in trace element analysis
Authors:B Maichin  P Kettisch  G Knapp
Institution:Institute for Analytical Chemistry, Micro- and Radiochemistry, Technical University Graz, Technikerstra?e 4, A-8010 Graz, Austria e-mail: knapp@analytchem.tu-graz.ac.at, AT
Anton Paar GmbH, Postfach 58, A-8054 Graz, Austria, AT
Abstract:Investigations of microwave assisted drying of sample materials and microwave assisted evaporation of aqueous sample solutions and acidic digestion residues were accomplished by means of special rotors for the microwave digestion system MULTIWAVE. To check the results obtained by microwave assisted drying, the samples were also conventionally dried at 105?°C in an oven. The following samples have been dried: 10 g each of meat, fish, apple, cucumber, potato, mustard, yogurt, clay and marl; 1 g each of certified reference material TORT 2 (lobster hepatopancreas), BCR 278 (mussel tissue) and BCR 422 (cod muscle); 500 g garden mould. Microwave assisted drying takes 40 min for organic samples and ¶30 min for inorganic material. Important is a slow increase of microwave power during the first 20 min. The results agree well with conventional drying at 105?°C. Losses of As, Se and Hg have been investigated for ¶3 CRMs. Only Se shows losses in the range of 20%. Losses of As, Be, Cd, Co, Cr, Cu, Fe, Hg, Li, Mg, Mn, Mo, Ni, Pb, Sb, Se, Sr, Ti, Tl, V and Zn after evaporation of aqueous samples and acidic solutions after wet digestion, respectively, have been investigated. 50 mL aqueous solution was evaporated almost to dryness within 25 min. The recovery of Hg is 40–50%, of Se 90–95% and of the other elements 97–102%. 0.2 g each of TORT 2, BCR 278 and BCR 422 have been digested with 4 mL nitric acid and 1 mL hydrochloric acid by means of the microwave digestion system MULTIWAVE. The digestion residue was evaporated almost to dryness and dissolved again in 10 mL diluted nitric acid. In this case no element losses have been observed. The measured concentration of As, Cd, Cu, Fe, Mn, Hg, Pb, Mo, Ni, Se, Sr, V and Zn agree very well with the certified values. An important prerequisite for good recoveries is not to evaporate the solutions to complete dryness.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号