首页 | 本学科首页   官方微博 | 高级检索  
     


Mass accuracy—TOF-SIMS
Authors:F.M. Green  I.S. Gilmore  M.P. Seah
Affiliation:Quality of Life Division, National Physical Laboratory, Teddington, Middlesex TW11 0LW, UK
Abstract:A study is presented of the factors affecting the calibration of the mass scale for time-of-flight SIMS (TOF-SIMS). The effect of the ion kinetic energy, emission angle and other instrumental operating parameters on the measured peak position are determined. This shows clearly why molecular and atomic ions have different relative peak positions and the need for an aperture to restrict ions at large emission angles. A calibration protocol is developed which gives a fractional mass accuracy of better than 10 ppm for masses up to 140 u. The effects of extrapolation beyond the calibration range are discussed and a recommended procedure is given to ensure that accurate mass is achieved within a selectable uncertainty for large molecules.
Keywords:Static SIMS   Accurate mass   Identification   Calibration   TOF-SIMS   G-SIMS
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号