Microcolumn liquid chromatography with thermionic detection of the enantiomers of O-ethyl S-2-diisopropylaminoethyl methylphosphonothioate (VX) |
| |
Authors: | Charles E Kientz Jan P Langenberg Udo A Th Brinkman |
| |
Abstract: | An improved interface for the on-line coupling of microcolumn liquid chromatography (micro-LC) with thermionic detection (TID) is described. Modifications have been made to enable separate adjustment of the eluent introduction and the detector flame temperature in order to improve the sensitivity and ease of use of the system. The micro-LC-TID was used for the chiral separation of the nerve agent O-ethyl S-2-diisopropylaminoethyl methylphosphonothioate (VX). Baseline separation for the enantiomers of VX was obtained on Chiralcel OD using 1% isopropanol in hexane as the eluent. The detection limit of VX using 60 nl injections is ca. 5 μg/ml (ppm range). However, when using large-volume injections (10 μl) the detection limit is ca. 25 ng/ml (ppb range). |
| |
Keywords: | Chemical warfare agents Microcolumn liquid chromatography Enantiomeric separation |
|