数字微镜近红外光谱仪光学系统设计与实验
引用本文:莫祥霞,温志渝,张智海,郭媛君.数字微镜近红外光谱仪光学系统设计与实验[J].光子学报,2014,40(9):1356-1360.
作者姓名:莫祥霞  温志渝  张智海  郭媛君
作者单位:(重庆大学 新型微纳器件与系统技术国家,光电技术及系统教育部重点实验室,重庆 400044)
基金项目:重庆市院士基金项目“基于MOEMS食品安全监测仪关键技术的研究”。;国家中央高校基本科研业务专项资金
摘    要:针对传统光谱仪体积大、成本高、检测速度慢、需样品前处理等不足,提出了利用数字微镜面阵(DMD)实现光谱谱面分割分时选通的近红外光谱仪光学系统.首先,对比传统光路介绍单探测器微型光谱仪系统测量原理|然后结合DMD特性提出光路方案,根据几何光学原理进行初步光学元件选型和光路结构设计,利用ZEMAX光学软件对光路进行仿真,确定结构参数|最后,搭建实验平台,进行光路测试.实验结果表明:该系统光路尺寸为70 mm×130 mm,测量波段为(900~1 500 nm),分辨率可达19 nm|在能量损失较小的情况下,减小狭缝尺寸可提高光学分辨率,狭缝的极限尺寸为200 μm|减小狭缝子午面高度可减小谱面内弯曲现象.本系统基本满足近红外分光,实现单点探测器光谱测量的要求.

关 键 词:近红外  数字微镜  光谱仪  单探测器
收稿时间:2011-04-07

Design and Experiment of Digital Mcro-mirror Spectrometer Optical System
MO Xiang-xia,WEN Zhi-yu,ZHANG Zhi-hai,GUO Yuan-jun.Design and Experiment of Digital Mcro-mirror Spectrometer Optical System[J].Acta Photonica Sinica,2014,40(9):1356-1360.
Authors:MO Xiang-xia  WEN Zhi-yu  ZHANG Zhi-hai  GUO Yuan-jun
Institution:(National Key Laboratory of Fundamental Science of Micro/Nano-Device and System Technology|
The Key Laboratory for Opto-eletronic Technology and System(Ministry of Education),Chongqing University,
Chongqing 400044,China)
Abstract:A compact,low cost,high speed,non-destructive testing NIR (near infrared) spectrometer optical system is developed based on digital micro-mirror device (DMD).The application of DMD as wavelength scanning element in a traditional lens optical system enables the design of compact grating spectrometers capable of acquiring full spectra using a single detector element.Firstly,comparing with the traditional optical system,there is a new structure with a single detector.With the characteristics of DMD,the structure of the spectrometer system is proposed.After calculating the parameters of the optical path,ZEMAX optical software is used to simulate the system.Finally,the prototype is fabricated and calibrated.Designed for a wavelength range between 900 nm~1 500 nm,the spectrometer optical system features a spectral resolution of 19nm with the area of 70 mm×130 mm.If the width of slit is more than 200 μm,decreasing its width can increase the resolution of this prototype and the change of intensity is slow.Adding an aperture in the prototype can reduce the curved of the slit image in the spectrum.The system satisfies the demand of Near Infrared(NIR) micro spectrometer with a single detector.
Keywords:   Near infrared  Micro-mirror  Spectrometer  Single detector
点击此处可从《光子学报》浏览原始摘要信息