首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A fast method to measure the thickness of removed layer from etching of SSNTD based on EDXRF
Authors:C W Y Yip  J P Y Ho  D Nikezic and K N Yu
Institution:

Department of Physics and Materials Science, City University of Hong Kong, Tat Chee Avenue, Kowloon Tong, Kowloon, Hong Kong

Abstract:Solid state nuclear track detectors are commonly used for measurements of concentrations of radon gas and/or radon progeny. All these measurements depend critically on the thickness of the removed layer during etching. However, the thickness of removed layer calculated using the etching period does not necessarily provide a sufficiently accurate measure of the thickness. For example, the bulk etch rate depends on the strength of stirring during etching for the LR 115 detector. We propose here to measure the thickness of the removed layer by using energy-dispersive X-ray fluorescence spectrometry. In the present work, a reference silver nitrate pellet is placed beneath the LR 115 detector, and the fluorescence X-ray intensity for silver is then measured. We have found a linear relationship between the X-ray intensity and the thickness of the removed layer for LR 115 detector. This provides a fast method to measure the thickness of removed layer from etching of LR 115 detector. However, this method was found to be inapplicable for the CR-39 detector. Therefore, alternative methods have yet to be explored for the CR-39 detector.
Keywords:SSNTD  Bulk etch  CR-39  LR 115  EDXRF