首页 | 本学科首页   官方微博 | 高级检索  
     


Parameterization of detector efficiency for the standardization of NAa with stable low flux reactors
Authors:G. Kennedy  J. St-Pierre
Affiliation:(1) Nuclear Engineering Institute, Ecole Polytechnique, Downtown, P.O. Box 6079, H3C 3A7 Montreal, Quebec, Canada
Abstract:With SLOWPOKE and MNS reactors which have reproducible neutron fluxes, the standardization of multielement NAA can be reduced to measuring activation constants once for all elements and then determining relative detection efficiencies for new detectors and counting geometries. In this work, a method has been developed for the parameterization of the efficiency of gemanium detectors. The gamma-ray detection efficiency was measured as a function of energy and distance for three detectors. The variation with distance was found to follow a modified EID law, within 1%, for point sources 1 mm to 250 mm from the detector. A model, including coincidence summing corrections, was developed to calculate efficiency for NAA samples; it requires 16 measured parameters. Tests showed that the calculated relative detection efficiencies are accurate to better than 3% for close counting geometries and sample volumes up to a few millilitres. Areas of possible improvement to the accurarcy of the method are suggested.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏