On the Use of ICP-MS for Measuring Plutonium in Urine
Authors:N Baglan  C Cossonnet  P Pitet  D Cavadore  L Exmelin  P Berard
Institution:(1) Institut de Protection et de Sûreté Nucléaire, Département de Protection de la Santé de l'Homme et de Dosimétrie, Service de Dosimétrie, IPSN, BP n° 6, F 92265 Fontenay-aux-Roses Cedex, France;(2) COGEMA, LAM Marcoule, BP170, F 30206 Bagnols sur Ceze Cedex, France;(3) COGEMA, LAM La Hague, F 50444 Beaumont Hague Cedex, France
Abstract:The analytical protocols currently used to measure plutonium in urinary excretion consist of radiochemical purification, the source preparation by electroplating and alpha spectrometry measurements. Such a procedure is of limited relevance to the individual monitoring of workers exposed to plutonium and mixed oxides. The use of ICP-MS, which takes much shorter counting time, was investigated with the aim of assessing the capabilities of this analytical tool for the determination of urinary excretion of plutonium. It has been shown that the detection level of 1 mBq·1-1 could be achieved after a radiochemical purification process for a standard ICP-MS set-up.