Design of a high-sensitivity negative ion source time-of-flight mass analyzer assembly created by cylindrical electrodes with a common axis
Authors:Ioanoviciu D  Cuna C  Cosma V  Albert I  Szilagyi Edina
Institution:National Institute for Research Development of Isotopic and Molecular Technologies, P.O. Box 700, Ro-3400 Cluj-Napoca 5, Romania. ioanovic@oc1.itim-cj.ro
Abstract:The new design incorporates the negative ion source and the mass analyzer, both constructed from cylindrical electrodes. The ion source is formed by three gridded cylindrical electrodes: a pulsed grid, the intermediate grid and the final accelerating grid. During a first time lapse, the electrons penetrate through the pulsed grid into the retarding field between this grid and the intermediate grid. The electrons are turning at some depth inside this intergrid space, where the attachment to neutral molecules most probably occurs. Next, the pulsed grid becoming strongly negative and ions are extracted towards the final acceleration grid. The ions from the cylindrical surface where they were created concentrate on the common axis of the electrodes (lateral focusing). The source lateral and time focus are coincident. A cylindrical electrostatic mirror is fitted to the source. The design, with a single stage, ensures also lateral focusing of the ions diverging from the common axis of the electrodes. The mirror electric and geometric parameters were selected to ensure both lateral and time focusing on the final detector with subsequent high luminosity. The basic parameters of the specific negative ion source time-of-flight mass analyzer design proposed here, are ion source final acceleration, intermediate, pulsed cylindrical grid radii 10, 20 and 30 mm, respectively, electrostatic mirror earthed grid and ion turning points surface radii 0.6 and 0.8 m, respectively. Ion packet smearing by the ion energy spread (resulting from the initial electron energy spread as electrons are turning at different depths inside the ionization region, from the moment when ions were created, being accelerated towards the pulsed grid during ionization) and by the turnaround time inside the cylindrical field was accounted for. Maintaining very high sensitivity, a resolution of the order of 100 is expected.
Keywords:Negative ions  time-of-flight analyzer  ion source  cylindrical reflectron  explosives
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号