首页 | 本学科首页   官方微博 | 高级检索  
     


Investigation of performance enhancement of flame photometric signal using a cylindrical impactor in the spray chamber
Authors:Kékedy-Nagy L
Affiliation:Department of Analytical Chemistry, Babe?-Bolyai University, Str. Arany J. 11, RO 3400 Cluj-Napoca, Romania.
Abstract:The effect of a cylindrical impactor in the spray chamber of a flame photometer on the characteristics of the aerosol formed, on the transport-efficiency of the nebulizer-spray chamber system, as well as on the analytical signal has been investigated. The variables were: the diameter Phi of the cylindrical impactor and its position, i.e., the distance d in front of the tip of pneumatic nebulizer capillary where the impactor was fixed. The performance characteristics of an 1 mm Phi impactor was fixed at d=5 mm was superior to that of a conventional 8 mm Phi spherical bead.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号