High performance liquid chromatographic determination of aluminium in natural waters in the form of its lumogallion chelate |
| |
Authors: | Zhou C Y Wu J Chi H Wong M K Koh L L Wee Y C |
| |
Institution: | Department of Chemistry, National University of Singapore, Lower Kent Ridge Road, Singapore 0511 Singapore. |
| |
Abstract: | A high performance liquid chromatographic method for the determination of ultra trace amount of aluminium in natural waters has been developed using lumogallion as a precolumn reagent for fluorimetric detection. The highly fluorescent Al-lumogallion chelate (lambda(ex) 500 nm, lambda(em) 574 nm) was separated on a LiChrosorb RP 18 column with an eluent containing 3:7 acetonitrile/0.02M potassium hydrogen phthalate buffer (pH 4.7) containing 10(-5)M lumogallion. The proposed system provides a simple, quick, selective and sensitive method for the determination of ultra-trace amount of aluminium in water samples. The detection limit defined as three times the standard deviation of the blank signal, was 0.05 mug/l. in water samples for 100 mul injection. The tolerance limits were 5 mg/l. for Fe(III) and F(-) and over 10 mg/l. for other foreign ions. The sensitivity of the method was independent of salinity. This method had been used for the direct determination of aluminium in both tap and coastal sea-waters without any preconcentration steps. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|