Abstract:A highly sensitive and selective voltammetric procedure is described for the determination of trace amounts of iron. The procedure is based on the adsorptive collection of an iron-thiocyanate-nitric oxide complex on a hanging mercury drop electrode. The adsorbed complex catalyzes the reduction of nitrite in solution, which gives a detection limit of 40 ppt iron (30 s accumulation). The stripping current increases linearly with iron concentration up to 80 ppb. The relative standard deviations are 4.2% and 1.6% at 0.5 ppb and 40 ppb respectively. Most of the common ions, except cobalt, do not interfere with the determination of iron. The procedure is applied to determine iron in biological samples, natural waters and analytical-grade chemicals.