首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Sequential injection wetting film extraction applied to the spectrophotometric determination of chromium(VI) and chromium(III) in water
Authors:Luo Y  Nakano S  Holman D A  Ruzicka J  Christian G D
Institution:Center for Process Analytical Chemistry, Department of Chemistry, Box 351700, University of Washington, Seattle, WA 98195, USA.
Abstract:The spectrophotometric determination of Cr(VI) and Cr(III) via sequential injection was used to demonstrate the sensitivity enhancement provided by a newly developed wetting film extraction system. The reaction product of Cr(VI) with 1,5-diphenylcarbazide was ion-paired with perchlorate and extracted into an organic wetting film consisting of octanol and 4-methyl-2-pentanone on the inner wall of a Teflon tube. The wetting film, with the extracted analyte, was then eluted with 100 mul acetonitrile and the analyte determined spectrophotometrically at 546 nm. Important optimized parameters were the selection of wetting film and elution solvents, the flow rate, the length and diameter of the extraction coil and the conditions for the formation of the ion paired chelate. Cr(III) was previously oxidized to Cr(VI) and calculated as the difference between total Cr and Cr(VI). An enrichment factor of 25 and a detection limit of 2.0 mug l(-1) Cr(VI) were achieved with a sampling frequency of 17 h(-1). The calibration curve was linear up to 100 mug l(-1) Cr(VI) (r = 0.999). The relative standard deviations were 2.8 and 2.0% at the 25 and 100 mug l(-1) levels.
Keywords: