Flow injection fluorimetric determination of thiourea |
| |
Authors: | Pérez-Ruiz T Martínez-Lozano C Tomás V Casajús R |
| |
Affiliation: | Department of Analytical Chemistry, Faculty of Chemistry, University of Murcia, Murcia, Spain. |
| |
Abstract: | A flow injection configuration for the fluorimetric determination of thiourea is proposed. The procedure is based on the rapid oxidation of thiourea by thallium(III) with concomitant formation of fluorescent thallium(I). Linear calibration graphs were obtained between 5 x 10(-7) and 1 x 10(-5)M, with a sampling rate of 90 samples/hr. The usefulness of the method was tested in the determination of thiourea in fruit juices and fruit peels. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|