首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Flow injection reagent introduction by supported liquid and nafion membranes: determination of phosphate
Authors:Chalk S J  Tyson J F
Institution:Department of Chemistry, University of Massachusetts, Amherst, MA 01003-0035, USA.
Abstract:The use of membrane tubing for the introduction of reagents for the determination of phosphate in waters by flow injection analysis was studied. The use of membranes eliminates the need for confluence points in the design of flow injection manifolds. This increases the sensitivity of the manifold by providing a sufficient reagent excess for the reaction without diluting the sample. Methods for the introduction of acid, molybdate and hydrazine were devised for the determination of phosphate by the Molybdenum Blue method. Several membranes were examined and Nafion and Accurel (microporous polypropylene) were found to be most useful. Molybdate introduction was achieved using a supported liquid membrane (SLM). Calibration was linear and a detection limit of 12 ppb phosphate (4 ppb phosphorus) was obtained.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏