首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A comparison of different HPLC detector designs based on the dropping mercury electrode principle
Authors:H B Hanekamp  P Bos  U A Th Brinkman and R W Frei
Institution:(1) Dept. of Analytical Chemistry, Free University, De Boelelaan 1083, 1081 HV Amsterdam, The Netherlands
Abstract:Summary Different designs of dropping-mercury electrode (DME) detectors for HPLC were tested and compared to solid-state and UV detectors. The time constants and the corresponding response volumes were measured. A qualitative explanation for the flow pattern of these cells was attempted. With nitrobenzene as a test substance the detection limits and reproducibilities of the best DME detectors were comparable to those of a good UV detector. The linear dynamic ranges and response times for optimally designed electroanalytical cells are better than for UV detectors.
Zusammenfassung Vergleiche zwischen diesen Detektoren sowie einem Festkörperelektroden und einem UV-Detektorentyp wurden angestellt. Zeitkonstanten und Detektionsvolumen wurden gemessen und eine Erklärung für die Flußprofile gegeben. Die mit Nitrobenzol gemessenen Detektionsgrenzen und Reproduzierbarkeiten der besten Typen sind vergleichbar mit denen von guten UV-Detektoren. Der dynamisch lineare Bereich und die Ansprechzeit sind für optimal gebaute elektro-analytische Detektoren deutlich besser als für UV-Detektoren.
Keywords:Chromatographie  HPLC  Vergleich von Detektoren auf Basis Hg-Tropfelektrode
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号