首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Folded resonator-dual polarization competition laser displacement sensor
Authors:Zhiguang Xu  Shulian Zhang  Wenhua Du  Yan Li  Lianqing Zhang  Jun Zhu
Institution:The State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instruments, Tsinghua University, Beijing 100084, China
Abstract:The existing so-called “dual polarization competition laser displacement sensor” applies a cat’s eye resonator. The resonator could be disarranged when the cat’s eye reflector (CER) swings slightly with the rod of the slide rail in measuring the procedure and droops due to its deadweight in long-time operation. This influences the stability of the sensor significantly. In this paper, a novel cat’s eye folded resonator was built, based on which our system named “folded resonator-dual polarization competition laser displacement sensor” was set up. It has a high resolution of λ/16 (39.6 nm) and large measurement range of up to 17 mm. The misalignment sensitivities of traditional cat’s eye resonator and the folded one are both analyzed; and comprehensive experiments are carried out, through which a conclusion can be drawn out that the maximal adjusting angle and the maximal cross displacement of the CER in folded resonator are both about 5 times of those in cat’s eye cavity. Furthermore, the measurement results from this new sensor prove its high stability.
Keywords:42  55  Lt
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏