首页 | 本学科首页   官方微博 | 高级检索  
     


Construction and evaluation of a miniature electron ion coincidence analyzer mounted on a conflat flange with an outer diameter of 114 mm.
Authors:Takuhiro Kakiuchi  Eiichi Kobayashi  Koji K Okudaira  Narihiko Fujita  Masatoshi Tanaka  Kazuhiko Mase
Affiliation:School of High Energy Accelerator Science, The Graduate University for Advanced Studies, 1-1 Oho, Tsukuba 305-0801, Japan. kakiuchi@post.kek.jp
Abstract:We have developed a miniature electron ion coincidence (EICO) analyzer mounted on a conflat flange with an outer diameter of 114 mm. It consists of a cylindrical mirror analyzer (CMA), a time-of-flight ion mass spectrometer (TOF-MS), a commercially available linear motion feed through, and a tilt adjustment mechanism. Each sample surface was irradiated by synchrotron radiation, and the energies of emitted electrons were analyzed and detected by the CMA, while desorbed ions were collected by the TOF-MS in coincidence with the electrons. The performance of the EICO analyzer was tested by measuring the Auger-electron H+ photoion coincidence spectrum of condensed water at 4a1 <-- O 1s resonance.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号