Abstract:An on-line gas diffusion/dialyser unit has been designed and incorporated into the manifold of a flow injection system for the simultaneous determination of two different species in a single sample injected into the FIA system. A fast and reliable fully automated two-component flow injection procedure has been carried out where a single sample (50 mgrl industrial effluent water) is injected into a carrier stream and simultaneously gas diffused and dialysed in a single on-line gas diffusion/dialyser unit for the determination of ammonia and chloride, respectively. The results obtained for the ammoniacal nitrogen and chloride in industrial effluent water compared well with data obtained by standard methods. The relative standard deviation for industrial effluent water samples with concentrations of ammoniacal nitrogen in the range between 0.050 and 1 g/l and chloride from 0.127 to 2.531 g/l have been better than 1.0% for ammoniacal nitrogen and 0.5% for chloride, respectively. The proposed system could be attractive for routine analyses of industrial effluent water.