Abstract: | A printed amperometric glucose sensor based on glucose oxidase adsorbed on crystals of tetrathiafulvalene-tetracyanoquinodimethane (TTF-TCNQ) is described. The sensitivity and the stability of the sensor are affected by the binder and solvent used for the preparation of the GOD.TTF-TCNQ paste. The sensors are continuously used in a flow injection analysis (FIA) system under continuous polarization at 0.15 V (vs Ag/AgCl) at 37°C. The developed sensors exhibit a large response current, an extended linear range and oxygen independence. The sensors can be used for more than 3 months. The GOD.TTF-TCNQ paste is suitable for the preparation of planar sensor by screen printing method. |