首页 | 本学科首页   官方微博 | 高级检索  
     


Continuous flow and flow injection stripping voltammetric determination of silver(I), mercury(II), and bismuth(III) at a bulk modified graphite tube electrode
Authors:Ruidong Ye  Soo Beng Khoo
Affiliation:Department of Chemistry, National University of Singapore, 10 Kent Ridge Crescent, Singapore 119260
Abstract:The epoxy-impregnated graphite tube electrode bulk modified with 2-mercaptobenzoxazole, employed in a wall-jet configuration, was found to be useful for the continuous flow and flow injection stripping voltammetric determinations of AgI, HgII and BiIII. For continuous flow, detection limits for AgI, HgII and BiIII were 1.8 × 10−10 M, 1.9 × 10−9 M and 9.5 × 10−9 M, respectively (10 min accumulation, S/N = 3). Precisions for 5.00 × 10−9 M AgI, 1.00 × 10−8 M HgII and 1.00 × 10−7 M BiIII were 10.5%, 5.77 % and 7.90% (relative standard deviations, n = 6), respectively. In the case of flow injection stripping, with a 500 μL injection loop, detection limits of 0.59 ng, 2.0 ng and 120 ng were obtained for AgI, HgII and BiIII, respectively (S/N = 3). Selected metal ions, inorganic and organic substances were investigated for interferences. The electrode was tested with a certified sample and then applied to the determinations of the metal ions in a urine and a sea-water sample.
Keywords:Stripping Voltammetry  Graphite-epoxy electrode  Silver  Mercury  Bismuth  Flow Analysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号