Development of a hypoxanthine biosensor based on immobilized xanthine oxidase chemically modified electrode
Authors:Shengshui Hu  Chung-Chiun Liu
Affiliation:Electronics Design Center & Edison Sensor Technology Center, Case Western Reserve University, Cleveland, OH 44106, USA

On leave from: Department of Chemistry, Wuhan University, P. R. China

Abstract:An amperometric biosensor for hypoxanthine determination has been developed. The sensor uses a Nafion-paraquat chemically modified glassy-carbon electrode. It detects the oxygen consumed by the enzymatic reaction catalyzed by xanthine oxidase which is immobilized on the electrode surface. The sensor responds linearly to hypoxanthine over the concentration range of 1 × 10−6 M −2 × 10−4 M. The biosensor can be reused for more than 100 times without significant deterioration in performance. After 32 days storage at 3–5°C, the sensor response remains at 68% of the initial level. The high sensitivity, selectivity and stability of this biosensor demonstrates its practical applicability for a simple, rapid and economical determination of hypoxanthine in fish samples.
Keywords:Biosensor  Enzyme electrode  Modified electrode  Amperometry  Hypoxanthine  Xanthine oxidase  Fish